You are viewing the site in preview mode
Skip to main content
|
Device
|
Sensitivity with 95% confidence interval
|
Specificity with 95% confidence interval
|
|---|
|
ProPex II
|
93% [68%;100%]
|
100% [78%; 100%]
|
|
Elements Apex Locator
|
93% [68%;100%]
|
100% [78%; 100%]
|
|
Apex NRG
|
100% [78%; 100%]
|
100% [78%; 100%]
|
|
RayPex 5
|
100% [78%; 100%]
|
100% [78%; 100%]
|
|
RayPex 6
|
93% [68%;100%]
|
100% [78%; 100%]
|
- The Apex NRG and RayPex 5 also detected all perforations (100% sensitivity; 95% confidence interval [78%; 100%]). The ProPex II, Elements Apex Locator, and RayPex 6 detected 14 of 15 perforations (93% sensitivity; 95% confidence interval [68%;100%]).